Vcheck Quality Control tests can be run if you suspect your system isn’t working correctly or to verify that you are performing the test correctly.

[When to use the control]
· Whenever your results are not in accordance with your expectations
· If you have repeated a test, and the result is still lower or higher than expected
· When you want to check the performance of the analyzer and test device
· If you have dropped the analyzer

[Before using the Control Kit]
· Check the expiration date of the Control Pouch. Do not use after the expiration date.