Coanda - 14 - Systems Control - Syringe Pumps

Опубликовано: 15 Июль 2026
на канале: Coanda Research & Development
75
1

The stepper motors that typically drive syringe pumps generally have 200 natural, magnetically-aligned, steps per revolution. Microsteps, made by energizing two motor windings at the same time, do not produce uniform step sizes. Here we’ll examine the consequences for syringe pump flow.

Your syringe pump specs should indicate the advance per microstep, and what the microsteps are. 1/128th steps are very common! With the syringe diameter, flow rate and advance per microstep, you can calculate how often the microsteps occur. At minimum flow rates, it’s frequently several seconds between microsteps and hundreds of seconds between full steps.

The combination of these uneven, spaced-out steps and the stiction of the syringe plunger on the syringe wall results in an intermittent liquid delivery, where the liquid flows for a short time then stops then starts again. The delivered flow rate may indeed still be as intended when averaged over a large timeframe, such as a minute, but the actual flow rate at any given moment fluctuates between zero and a much higher throughput. This can be a particularly problematic situation when micro dosing fluids into a process stream such as adding a coagulant to a mixture or applying adhesive to a surface.

If you’re running your syringe pump at the low end of its speed range, find a capillary tube, connect it to the pump and watch the meniscus advance. This might not sound exciting, but if you care about the reliability of your flow rate, you might end up quite excited.