Syneron DS Hand Piece Repair
P4L 52555, S/N F18340962
Error: Needs Service/Rebuild
Service/Labor: Complete rebuild
Replace Lamp and seal kit
Reset pulse count to zero
Test and check cooling system
Polish Crystal
Calibration and check output energy.