Biomedical Medtronic Clinical Specialist Checking Heart Parameters Using Pacemaker Programmer

Опубликовано: 15 Июль 2026
на канале: A2 Tech
239
1

Associate Clinical Specialist having Biomedical Engineering Background goes for Pacemaker Technical Implant Support.. pacemaker Implants requires Proper Positioning of Pacemaker leads..
If Dual chamber is there, than it contains 2 leads or wire , which needs to be placed in Right Atrium and Right Ventricle By Cardiac Surgeon-Electrophysiologist While seeing Fluoroscopic view..the Parameters when lead is positioned Into Proper Anatomical Positions like Septum or Apex for right ventricle.. By seeing Into Programmer which is connected with Pacemaker leads using Testing cable..Parameters like Current of Injury by looking into IntraCardiac Electrogram, Sensing , Pacing - Threshold,Lead Impedance etc are taken into Consideration .
Than after Programming Base rate and other Algorithms like AutoCapture and etc Done while and Ecg is taken in order to make sure that Programmed Parameters are working fine or not..